      Model: 732C and 734C DC Voltage Reference Standards

      The simple way to maintain and disseminate the volt. The Fluke Calibration 732C DC Voltage Reference Standard is a RoHS compliant direct voltage reference used to maintain the volt in primary and secondary standards laboratories.

      Individual 732C voltage standards provide 10 V, 1 V, and 0.1 V outputs and may be transported easily to remote locations while the dc reference is maintained in the laboratory.

      The Fluke 734C DC Voltage Reference Standard consists of four electrically and mechanically independent 732C DC Reference Standards and a rack-width enclosure. Two model families, Base and Select, differ in that the Select models are two times more stable than the Base models. We Are the Authorize distributor in Bangaladesh and grateful to Tools Bangladesh.

      Fluke PTi120 Pocket Thermal Camera

      Key Features:

      Get the infrared troubleshooting power you need.

      • Automatically organize and file thermal images with Fluke Connect Asset Tagging
      • Fully radiometric thermal imager gives you measurement data to the pixel with 120 x 90 infrared resolution (10,800 pixels)
      • 5” LCD touchscreen display for easy troubleshooting
      • -20 °C to +150 °C temperature measurement range
      • Six color palettes
      • Can withstand drop up to 1 meter
      • IP 54 enclosure rating

      Specifications: 732C and 734C DC Voltage Reference Standards

      Line voltage is accepted as shown in Table 1. AC line current at 120V ac is typically 0.13A. Table 1. Mains

      732C Line Voltage Setting Line Voltage Accepted Frequency Accepted
      100 V 90 V to 110 V 50 Hz/60 Hz
      120 V 108 V to 132 V 50 Hz/60 Hz
      220 V 198 V to 242 V 50 Hz/60 Hz
      240 V 216 V to 264 V 50 Hz/60 Hz
      Battery Battery Operation At full charge, the internal battery operates the Product for a minimum of 72hours at 23±5 °C, with 0 mA to 0.1 mA total current drain at the outputs.
      Charging Time <36 hours with self-contained automatic battery charger
      External DC Input Rear-panel input for external 12 V dc to 15 V dc powers the Product indefinitely. The dc source must be rated ≥300 mA.
      Isolation Resistance from any of the Product binding posts to earth (chassis) ground or to ac line power is >10 000 MΩ shunted by <1000 pF.
      Guard and Ground Terminals Chassis ground connections are provided on both the front and rear panels. Access to the internal guard is provided by a front-panel binding post.
      Output Protection All outputs can be shorted indefinitely without damage to the Product. The 10 V output can withstand voltages from other sources as follows:

       

      1. For voltages ≤220 V dc, the Product is protected for a maximum of 50 mA continuous current.
      2. For voltage ≤1100 V dc, the Product is protected for a maximum of 25 mA continuous current or up to 0.6 joules for short periods of time.
      Environmental
      Specified Operation Temperature Range: 15 °C to 35 °C
      Relative Humidity: 0% to 90% to 28 °C, to 80% to 35 °C, and to 50% to 50 °C, Noncondensing
      Altitude: 0 m to 1830 m (0 ft to 6000 ft)
      Non-specified Operation Temperature Range: 0 °C to 50 °C
      Relative Humidity: 0% to 90% Noncondensing
      Altitude: 0 m to 3050 m (0ft to 10 000ft)
      Storage (Battery Removed) Temperature Range: -40 °C to 50 °C
      Relative Humidity: Noncondensing
      Altitude: 0m to 12 200 m (0 ft to 40 000 ft)
      Electromagnetic Compatibility

      (EMC) The Product operates in Standards Laboratory environments where the radio frequency (RF) environment is highly controlled. 

      International

      IEC 61326-2-1; CISPR 11: Group 1, Class A

      Controlled Electromagnetic Environment

      Group 1 equipment has intentionally generated and/or use conductively coupled radio-frequency energy which is necessary for the internal functioning of the equipment itself.

      Class A equipment is equipment suitable for use in all establishments other than domestic and those directly connected to a low voltage power supply network which supplies buildings used for domestic purposes.

      Emissions which exceed the levels required by CISPR 11 can occur when the equipment is connected to a test object. The equipment may not meet the immunity requirements of 61326-1 when test leads and/or test probes are connected.

      USA (FCC) 47 CFR 15 subpart B, this product is considered an exempt device per clause 15.103
      Korea (KCC)

      Class A Equipment (Industrial Broadcasting & Communication Equipment)

      This product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and not to be used in homes.

      Safety
      Safety IEC 61010-1, Installation Category II, Pollution degree 2
      Ingress Protection IEC 60529: IP20
      Mechanical Specifications
        734C Dimensions 732C and 732C-7001 Dimensions
      Height 17.8 cm (7.0 in) 13.4 cm (5.28 in)
      Width 43.2 cm (17.0 in) 9.8 cm (3.85 in)
      Depth 50.3 cm (19.8 in) including handles 40.6 cm (16.0 in)
      Weight 30.4 kg (67 lb) 5.91 kg (13 lb)

      Performance Specifications

      Output Voltages

      10 V, 1 V, and 0.1 V are provided at separate binding posts referenced to the VCOM binding post.

      Stability

      Stability for the 732C outputs at Tcal ±1 °C and IN CAL indicator on is specified in Table 2.

      Table 2. Standard Stability

      Output Voltage Stability (± µV/V)
      30 Days 90 Days 1 Year
      10 V 0.3 0.8 2.0
      1 V 0.6 1.2 3.0
      0.1 V 1.2 2.9 9.8

      Table 3. Select Stability

      Output Voltage Stability (± µV/V)
      30 Days 90 Days 1 Year
      10 V 0.3 0.8 1.0
      1 V 0.6 1.2 2.5
      0.1 V 1.2 2.9 8.0

      Noise at the Output Terminals

      Output noise is specified for both day-to-day observations and for short-term observations, at k=1.

      Table 4. Noise at the Output Terminals

      Output Voltage S1 (±µV/V)

      732C and 734C DC Voltage Reference Standards

      Includes:

      • 732C and 734C DC Voltage Reference Standards

       

      Accessories: 732C and 734C DC Voltage Reference Standards

      Model Name Description
      732C

      10V DC Reference Standard 

      732C/C

      10V DC Ref Std + Char   (ships hot – INTL)

      732C/S/C

      Select 10 V dc Ref Std + Characterization (ships hot – INTL)

      734C

      Base Model 4 Cell DC Reference Standard (ships cold)

      734C/04

      4 Cell DC Ref Std + 4 w/Char (ships hot – INTL)

      734C/S/04

      Select Model 4 Cell DC Reference Standard + 4 w/Char (ships hot) 

      Accessories common to all models:

      Accessory Description
      732C-7001

      External Battery and Charger

      732C-7002

      Transit Case, for qty. 1 or 2 732Bs or 732Cs

      734C-7001

      Instrument Enclosure

      Y734

      Rack-Mount Kit for 734A or 734A-700

      Based on 0 reviews

      0.00 Overall
      0%
      0%
      0%
      0%
      0%
      Be the first to review “732C and 734C DC Voltage Reference Standards”

      Your email address will not be published. Required fields are marked *

      Reviews

      There are no reviews yet.

      Close
      Close My Cart
      Close Wishlist
      Recently Viewed Close
      Close

      Close
      Navigation
      Categories